Comparison of the hybrid locking plate, standard dynamic compression plate, and standard dynamic compression plate augmented with bone cement for fixation of osteoporotic humeral shaft fractures : a cadaveric biomechanical study / Irewin A. Tabu [and two others]
Subject(s): Online resources: Abstract: BACKGROUND: Studies comparing the relative strength of polymethylmethacrylate (PMMA) augmented fixation, standard plating and locked compression plate (LCP) system are few. The use of either the bone cement-augmented dynamic compression plate or the Hybrid LCP constructs may provide an additional tool for the treatment of fractures in patients with osteoporosis. METHODS: Eighteen (18) osteoporotic cadaveric humeral bones were assigned randomly to each of three groups (Dynamic Compression Plate [DCP], DCP augmented with bone cement, and the Hybrid LCP system) and tested in anterior-posterior bending and torsion/external rotation. The load to failure values were obtained and the results for each specimen compared. RESULTS: Significant differences were observed between the standard DCP and Hybrid LCP group (p-value=0.012), and in the cement-augmented and Hybrid LCP group (p-value=0.099) in torsion/external rotation loading. No significant difference was observed between the standard DCP and bone-cement augmented group (p-value=0.248). No significant difference was observed among the three groups in terms of stiffness (p-value=0.3868) in the four-point anterior-posterior bending modality. Screw pull-out of the implant was observed only in the regular DCP group in torsion/external rotation loading stress. CONCLUSION: Significant differences were seen between the three constructs in torsion/external rotation but not in anterior-posterior four-point bending. Bone failure, but not screw pull-out, was seen in the Hybrid LCP and bone cement-augmented DCP groups in torsion. This study showed that the LCP system and the bone cement-augmented constructs may provide greater screw purchase to the osteoporotic humerus.| Cover image | Item type | Current library | Home library | Collection | Shelving location | Call number | Materials specified | Vol info | URL | Copy number | Status | Notes | Date due | Barcode | Item holds | Item hold queue priority | Course reserves | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Journal Article
|
DOH Central Library Electronic Resource Section | J000078 (Browse shelf(Opens below)) | Available | D0001J000078 |
Browsing DOH Central Library shelves,Shelving location: Electronic Resource Section Close shelf browser (Hides shelf browser)
BACKGROUND: Studies comparing the relative strength of polymethylmethacrylate (PMMA) augmented fixation, standard plating and locked compression plate (LCP) system are few. The use of either the bone cement-augmented dynamic compression plate or the Hybrid LCP constructs may provide an additional tool for the treatment of fractures in patients with osteoporosis.
METHODS: Eighteen (18) osteoporotic cadaveric humeral bones were assigned randomly to each of three groups (Dynamic Compression Plate [DCP], DCP augmented with bone cement, and the Hybrid LCP system) and tested in anterior-posterior bending and torsion/external rotation. The load to failure values were obtained and the results for each specimen compared.
RESULTS: Significant differences were observed between the standard DCP and Hybrid LCP group (p-value=0.012), and in the cement-augmented and Hybrid LCP group (p-value=0.099) in torsion/external rotation loading. No significant difference was observed between the standard DCP and bone-cement augmented group (p-value=0.248). No significant difference was observed among the three groups in terms of stiffness (p-value=0.3868) in the four-point anterior-posterior bending modality. Screw pull-out of the implant was observed only in the regular DCP group in torsion/external rotation loading stress.
CONCLUSION: Significant differences were seen between the three constructs in torsion/external rotation but not in anterior-posterior four-point bending. Bone failure, but not screw pull-out, was seen in the Hybrid LCP and bone cement-augmented DCP groups in torsion. This study showed that the LCP system and the bone cement-augmented constructs may provide greater screw purchase to the osteoporotic humerus.
In: Acta Medica Philippina, 2012 vol 46 (2) pages: 24-31